00:00:10,480 --> 00:00:11,800
Hello and welcome to the
00:00:11,800 --> 00:00:14,560
Somno touch home sleep
study instruction video.
00:00:15,320 --> 00:00:17,560
You will receive two
separate blue bags,
00:00:17,560 --> 00:00:20,120
one small and one slightly bigger.
00:00:21,720 --> 00:00:24,040
Included in the smaller bag is a small
00:00:24,040 --> 00:00:28,240
Somno touch recorder, which records
all aspects of the sleep study,
00:00:28,560 --> 00:00:32,200
and the blue measurement adapter
for the chest and abdominal bands.
00:00:33,200 --> 00:00:36,200
You will notice that there
are some disposable items.
00:00:36,680 --> 00:00:38,960
The SpO2 for the finger.
00:00:40,720 --> 00:00:42,600
The elastic measurement band
00:00:42,600 --> 00:00:44,480
and the nasal cannula.
00:00:49,920 --> 00:00:52,320
The other bag contains
the video camera,
00:00:52,320 --> 00:00:55,720
a tripod and a cable to connect
it to the power socket.
00:00:57,520 --> 00:00:59,320
Firstly, carefully attach a
00:00:59,320 --> 00:01:01,760
tripod to the base of the camera.
00:01:02,520 --> 00:01:06,960
Find a suitable position in the bedroom
to place the video camera, preferably
00:01:06,960 --> 00:01:10,800
close enough to see your child's head
and face, but also the body.
00:01:20,560 --> 00:01:22,800
Connect the camera to a power socket.
00:01:22,800 --> 00:01:25,400
Use an extension lead if necessary.
00:01:29,960 --> 00:01:32,280
Check to make sure the
little light located
00:01:32,280 --> 00:01:34,440
at the back of the camera turns on.
00:01:35,080 --> 00:01:37,040
When the recording time begins.
00:01:37,040 --> 00:01:40,600
You should see lights illuminate
on the front of the camera, indicating
00:01:40,600 --> 00:01:42,640
the start of the recording.
00:01:44,160 --> 00:01:46,120
Now lay out the bands.
00:01:46,840 --> 00:01:49,840
You will find two sets
of disposable bands,
00:01:50,160 --> 00:01:53,160
either small and medium
or medium and large.
00:01:53,880 --> 00:01:56,640
It is essential that the bands fit snugly.
00:01:57,760 --> 00:01:59,800
Attach one end of the band poppers
00:01:59,800 --> 00:02:02,400
to the blue measurement adapter.
00:02:04,360 --> 00:02:07,360
Thread the Somno touch to the chest band.
00:02:14,920 --> 00:02:17,240
Position the band on the upper chest,
00:02:17,240 --> 00:02:20,120
ideally over a vest for maximum comfort,
00:02:20,120 --> 00:02:22,040
then popper shut the band.
00:02:29,120 --> 00:02:32,520
Attach a lower band at the level of
the belly button and pop up.
00:02:35,040 --> 00:02:37,400
Now adjust both bands to fit,
00:02:37,400 --> 00:02:40,800
so that they feel secure,
but not too uncomfortably tight.
00:02:42,640 --> 00:02:46,480
Ideally, make these adjustments
whilst your child is standing up
00:02:46,480 --> 00:02:50,400
or laying down flat,
not whilst the child is sitting.
00:02:50,960 --> 00:02:53,640
They will be slack when laying down.
00:02:54,480 --> 00:02:56,880
Never attach a band
straight over the skin.
00:02:56,880 --> 00:03:00,360
The child needs to wear
a preferably tight top or pajama.
00:03:02,760 --> 00:03:05,760
Thread the cable under
the pajamas to the adapter.
00:03:06,680 --> 00:03:09,840
Plug the blue measurement
adapter to the some touch.
00:03:11,080 --> 00:03:12,920
It is really important that the blue
00:03:12,920 --> 00:03:15,920
on/off buttons on the right hand side.
00:03:16,320 --> 00:03:20,160
Otherwise, this can easily be
put on upside down by mistake.
00:03:21,800 --> 00:03:23,080
Nasal prongs.
00:03:24,040 --> 00:03:27,040
These measure the breathing
and snoring from the nose.
00:03:27,600 --> 00:03:30,800
The prongs need to be positioned
at the base of the nostrils,
00:03:30,800 --> 00:03:34,320
with the tubing draped over the ears
and secured under the chin.
00:03:34,960 --> 00:03:38,200
Place some tape across the
tubing on the cheeks to secure.
00:03:39,960 --> 00:03:42,320
Thread the tubing under
the pajamas and twist
00:03:42,320 --> 00:03:45,320
into the bottom of the
Somno touch control panel.
00:03:52,720 --> 00:03:55,400
If your child feels the
prongs are uncomfortable,
00:03:55,400 --> 00:03:58,240
you can remove them
and try to reapply them
00:03:58,240 --> 00:04:00,480
after your child goes to sleep.
00:04:00,480 --> 00:04:02,400
Attempt that 2 or 3 times
00:04:02,400 --> 00:04:07,360
and if your child still cannot settle,
it is okay to not use the nasal prongs.
00:04:11,520 --> 00:04:15,800
The SpO2 attachment measures
the oxygen in the blood and the heart rate.
00:04:16,400 --> 00:04:19,800
Attache this to one of your child's
fingers and feed the cable up
00:04:19,800 --> 00:04:21,120
through the sleeve.
00:04:45,040 --> 00:04:48,040
Plug the SpO2 into the Somno Touch.
00:04:49,320 --> 00:04:51,440
The setting up process is now complete.